Kodak LED II 20R Operator Guide

Calibration Troubleshooting

Diagnostics and Troubleshooting
5-14
September 2000
Calibration Troubleshooting
This section describes the problems that you may occasionally have when using 
the KODAK Device Calibration Software or the Calibration Software for the 
KODAK PROFESSIONAL LED II Printer. It also identifies the probable causes for 
these problems and provides solutions for correcting these problems.
Calibration Graph
The graph below is an indication that the calibration performed was out of 
tolerance.
This may be result from invalid densitometer readings. It is necessary to perform 
another calibration cycle if you see a graph that looks like the one above.
Numbered Error Messages
Error 
Code
Error Code /Status 
Message
Possible Cause/Subsystem
Possible Solution
1
The following file is 
locked:_filename_.
The system is trying to access a 
file that is currently in use.
Close any other applications that 
may be accessing the file.
2
The following file is 
protected:_filename_.
The system is trying to access a 
file that is protected.
Change the protections on the file 
to grant access to the user.
3
The following file was not 
found:_filename_.
The system was trying to access 
a file that it could not find.
Verify that the filename on one of 
the configuration dialog windows is 
correct.
4
Error:_oserror_occurred 
during operation:_operation_ 
on file:_filename_.
An operating system error 
occurred.
• Rerun the application.
• Reboot the operating system
5
The following file is in an 
incorrect format:_filename_.
The file does not conform to the 
application specification.
Verify the file format.